Ptx
SCPI Commands :
FETCh:BLUetooth:MEASurement<Instance>:MEValuation:TRACe:SGACp[:PTX]
READ:BLUetooth:MEASurement<Instance>:MEValuation:TRACe:SGACp[:PTX]
- class PtxCls[source]
Ptx commands group definition. 2 total commands, 0 Subgroups, 2 group commands
- fetch() List[float] [source]
# SCPI: FETCh:BLUetooth:MEASurement<Instance>:MEValuation:TRACe:SGACp[:PTX] value: List[float] = driver.multiEval.trace.sgacp.ptx.fetch()
Returns the values of the ‘Spectrum Gated ACP’ tables for EDR packets. The R&S CMW measures the adjacent channel power values PTX(f) in line with Bluetooth test specification.
INTRO_CMD_HELP: The number of valid results depends on the ACP measurement mode (method RsCmwBluetoothMeas.Configure.MultiEval.Sacp.Brate.Measurement.mode CH21 | CH79) :
If ‘ACP +/- 10 Channels’ is selected, the first 21 values contain the results for the relative channels –10, …, 0, …, +10; the remaining 58 values are not displayed.
If ‘ACP 79 Channels’ is selected, valid ACP values are available for all channels in the Bluetooth regulatory range.
Use RsCmwBluetoothMeas.reliability.last_value to read the updated reliability indicator.
- return:
acp: float 79 ACP results Range: -99.99 dBm to 99.99 dBm, Unit: dBm
- read() List[float] [source]
# SCPI: READ:BLUetooth:MEASurement<Instance>:MEValuation:TRACe:SGACp[:PTX] value: List[float] = driver.multiEval.trace.sgacp.ptx.read()
Returns the values of the ‘Spectrum Gated ACP’ tables for EDR packets. The R&S CMW measures the adjacent channel power values PTX(f) in line with Bluetooth test specification.
INTRO_CMD_HELP: The number of valid results depends on the ACP measurement mode (method RsCmwBluetoothMeas.Configure.MultiEval.Sacp.Brate.Measurement.mode CH21 | CH79) :
If ‘ACP +/- 10 Channels’ is selected, the first 21 values contain the results for the relative channels –10, …, 0, …, +10; the remaining 58 values are not displayed.
If ‘ACP 79 Channels’ is selected, valid ACP values are available for all channels in the Bluetooth regulatory range.
Use RsCmwBluetoothMeas.reliability.last_value to read the updated reliability indicator.
- return:
acp: float 79 ACP results Range: -99.99 dBm to 99.99 dBm, Unit: dBm