Current

SCPI Commands

CALCulate:BLUetooth:MEASurement<Instance>:MEValuation:PVTime:NMODe:CLASsic:CURRent
READ:BLUetooth:MEASurement<Instance>:MEValuation:PVTime:NMODe:CLASsic:CURRent
FETCh:BLUetooth:MEASurement<Instance>:MEValuation:PVTime:NMODe:CLASsic:CURRent
class Current[source]

Current commands group definition. 3 total commands, 0 Sub-groups, 3 group commands

class CalculateStruct[source]

Response structure. Fields:

  • Reliability: int: decimal See ‘Reliability Indicator’

  • Out_Of_Tol: float: float | ON | OFF Out of tolerance result, i.e. percentage of measurement intervals of the statistic count ([CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:SCOunt:PVTime CMDLINK]) exceeding the specified limits, see [CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:LIMit:BRATe:PVTime CMDLINK] and [CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:LIMit:EDRate:PVTime CMDLINK]. Range: 0 % to 100 %, Unit: %

  • Nominal_Power: float: float Average power during the carrier-on state Range: -99.99 dBm to 99.99 dBm, Unit: dBm

class FetchStruct[source]

Response structure. Fields:

  • Reliability: int: decimal See ‘Reliability Indicator’

  • Out_Of_Tol: float: float | ON | OFF Out of tolerance result, i.e. percentage of measurement intervals of the statistic count ([CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:SCOunt:PVTime CMDLINK]) exceeding the specified limits, see [CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:LIMit:BRATe:PVTime CMDLINK] and [CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:LIMit:EDRate:PVTime CMDLINK]. Range: 0 % to 100 %, Unit: %

  • Nominal_Power: float: float Average power during the carrier-on state Range: -99.99 dBm to 99.99 dBm, Unit: dBm

class ReadStruct[source]

Response structure. Fields:

  • Reliability: int: decimal See ‘Reliability Indicator’

  • Out_Of_Tol: float or bool: float | ON | OFF Out of tolerance result, i.e. percentage of measurement intervals of the statistic count ([CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:SCOunt:PVTime CMDLINK]) exceeding the specified limits, see [CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:LIMit:BRATe:PVTime CMDLINK] and [CMDLINK: CONFigure:BLUetooth:MEASi:MEValuation:LIMit:EDRate:PVTime CMDLINK]. Range: 0 % to 100 %, Unit: %

  • Nominal_Power: float: float Average power during the carrier-on state Range: -99.99 dBm to 99.99 dBm, Unit: dBm

calculate()CalculateStruct[source]
# SCPI: CALCulate:BLUetooth:MEASurement<Instance>:MEValuation:PVTime:NMODe:CLASsic:CURRent
value: CalculateStruct = driver.multiEval.powerVsTime.nmode.classic.current.calculate()

Returns the current, average, min, and max power results for BR/EDR, normal mode. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.

return

structure: for return value, see the help for CalculateStruct structure arguments.

fetch()FetchStruct[source]
# SCPI: FETCh:BLUetooth:MEASurement<Instance>:MEValuation:PVTime:NMODe:CLASsic:CURRent
value: FetchStruct = driver.multiEval.powerVsTime.nmode.classic.current.fetch()

Returns the current, average, min, and max power results for BR/EDR, normal mode. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.

return

structure: for return value, see the help for FetchStruct structure arguments.

read()ReadStruct[source]
# SCPI: READ:BLUetooth:MEASurement<Instance>:MEValuation:PVTime:NMODe:CLASsic:CURRent
value: ReadStruct = driver.multiEval.powerVsTime.nmode.classic.current.read()

Returns the current, average, min, and max power results for BR/EDR, normal mode. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.

return

structure: for return value, see the help for ReadStruct structure arguments.