SoBw
SCPI Command :
CONFigure:BLUetooth:MEASurement<Instance>:MEValuation:LIST:SEGMent<nr>:SCOunt:SOBW
- class SoBwCls[source]
SoBw commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(segment=Segment.Default) int [source]
# SCPI: CONFigure:BLUetooth:MEASurement<Instance>:MEValuation:LIST:SEGMent<nr>:SCOunt:SOBW value: int = driver.configure.multiEval.listPy.segment.scount.soBw.get(segment = repcap.Segment.Default)
Defines the statistic count in the segment. The last mnemonic denotes the measurement type: statistical modulation measurement, statistical power measurement, spectrum ACP measurement (BR, LE) , spectrum gated ACP measurement (EDR) , spectrum 20 dB bandwidth (occupied bandwidth) measurement (BR) .
- param segment:
optional repeated capability selector. Default value: S1 (settable in the interface ‘Segment’)
- return:
spec_obw_stat_cnt: numeric Statistic count Range: 1 to 1000
- set(spec_obw_stat_cnt: int, segment=Segment.Default) None [source]
# SCPI: CONFigure:BLUetooth:MEASurement<Instance>:MEValuation:LIST:SEGMent<nr>:SCOunt:SOBW driver.configure.multiEval.listPy.segment.scount.soBw.set(spec_obw_stat_cnt = 1, segment = repcap.Segment.Default)
Defines the statistic count in the segment. The last mnemonic denotes the measurement type: statistical modulation measurement, statistical power measurement, spectrum ACP measurement (BR, LE) , spectrum gated ACP measurement (EDR) , spectrum 20 dB bandwidth (occupied bandwidth) measurement (BR) .
- param spec_obw_stat_cnt:
numeric Statistic count Range: 1 to 1000
- param segment: