Le2M
SCPI Commands :
READ:BLUetooth:MEASurement<Instance>:MEValuation:SACP:LENergy:LE2M
FETCh:BLUetooth:MEASurement<Instance>:MEValuation:SACP:LENergy:LE2M
CALCulate:BLUetooth:MEASurement<Instance>:MEValuation:SACP:LENergy:LE2M
- class Le2MCls[source]
Le2M commands group definition. 3 total commands, 0 Subgroups, 3 group commands
- class CalculateStruct[source]
Response structure. Fields:
Reliability: int: decimal ‘Reliability indicator’
Nominal_Power: float or bool: float Average power during the carrier-on state Range: -99.99 dBm to 99.99 dBm
No_Of_Exceptions: float or bool: decimal Number of exceptions (channels ±3, ±4 … with an ACP above the ‘Exception PTx’ threshold ) Range: 0 to 99
- class ResultData[source]
Response structure. Fields:
Reliability: int: decimal ‘Reliability indicator’
Nominal_Power: float: float Average power during the carrier-on state Range: -99.99 dBm to 99.99 dBm
No_Of_Exceptions: int: decimal Number of exceptions (channels ±3, ±4 … with an ACP above the ‘Exception PTx’ threshold ) Range: 0 to 99
- calculate() CalculateStruct [source]
# SCPI: CALCulate:BLUetooth:MEASurement<Instance>:MEValuation:SACP:LENergy:LE2M value: CalculateStruct = driver.multiEval.sacp.lowEnergy.le2M.calculate()
Returns the ‘Spectrum ACP’ results for LE 2M PHY (…:LE2M…) and LE coded PHY (…:LRANge…) . See ‘View Spectrum ACP’. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.
- return:
structure: for return value, see the help for CalculateStruct structure arguments.
- fetch() ResultData [source]
# SCPI: FETCh:BLUetooth:MEASurement<Instance>:MEValuation:SACP:LENergy:LE2M value: ResultData = driver.multiEval.sacp.lowEnergy.le2M.fetch()
Returns the ‘Spectrum ACP’ results for LE 2M PHY (…:LE2M…) and LE coded PHY (…:LRANge…) . See ‘View Spectrum ACP’. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.
- return:
structure: for return value, see the help for ResultData structure arguments.
- read() ResultData [source]
# SCPI: READ:BLUetooth:MEASurement<Instance>:MEValuation:SACP:LENergy:LE2M value: ResultData = driver.multiEval.sacp.lowEnergy.le2M.read()
Returns the ‘Spectrum ACP’ results for LE 2M PHY (…:LE2M…) and LE coded PHY (…:LRANge…) . See ‘View Spectrum ACP’. The values described below are returned by FETCh and READ commands. CALCulate commands return limit check results instead, one value for each result listed below.
- return:
structure: for return value, see the help for ResultData structure arguments.